Journal of Cancer Sciences

Download PDF
Research Article

Breast Cancer Classification: A CAD System for a Combined Use of Elastography and B-Mode Sonography

Marcomini KD1*, Fleury EFC2 and H. Schiabel2

1Department of Electrical and Computer Engineering, University of São Paulo, Brazil
2Department of Electrical and Computer Engineering, University of So Paulo,Brazil
*Address for Correspondence: Marcomini KD, Department of Electrical and Computer Engineering, University of São Paulo, 400 Avenida Trabalhador São-Carlense, 13566-590 São Carlos, SP, Brazil
Submission: 05 August, 2020; Accepted: 12 September, 2020; Published: 20 November, 2020
Copyright: © 2020 Marcomini KD, et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Purpose: The aim of this study was to evaluate and compare the diagnostic performance of elastography, B-mode Ultrasound (US), and a combination of elastography and B-mode US for the differentiation between benign and malignant lesions.
Methods: A prospective study was carried out from July to December 2015, which 87 patients with 83 lesions were examined with conventional B-mode ultrasound and strain elastography. All the lesions had been proven by biopsy, resulting in 31 malignant and 52 benign lesions. A radiologist with 16 years of experience classified visually these cases. We also used a CAD sys-tem to classify the lesions classified visually by the most experienced radiologist and using a CAD system. The data obtained were compared with the results provided by another radiologist and a resident with 2 years of experience. Sensitivity, specificity and AUC for the three observers using the CAD system were calculated.
Results: The developed CADx system provided a diagnostic concordance in the classification of breast lesions from the different ways of contour determination (manual and automatic), allowing to reduce the diagnostic variability. In addition, the CADx system showed superior results to the visual analysis of the radiologist. When the radiologist associated both examinations (B-mode ultrasound and elastography), his visual analysis provided 87.1%, 55.8% and 0.714 of sensitivity, specificity and AUC, respectively. When we considered the result provided by the association between B-mode ultrasound and elastography images, the CADx system provided a comparative increase of about 7% of sensitivity and 17.2% of specificity, using the contour delimited by the most experienced radiologist. In addition, a positive influence was observed in the use of the computational tool by radiologists, since, on average, their sensitivity and specificity indexes also increased in relation to the conventional analysis, from 87.1% and 55.8% to 90.3% and 73.1%, respectively.
Conclusion: Thus, it can be concluded that the developed CADx system performed well in distinguishing benign from malignant lesions for both B-mode ultrasound and elastography. The AUC obtained was higher than the radiologist’s visual analysis in most of the cases analyzed.

Keywords

Breast cancer; Elastography imaging; Computer-Aided Diagnosis; Color map; Interobserver agreement

Introduction

Breast Ultrasound (US) is an important complementary technique for screening and has been proven to be useful in differentiating benign from malignant masses, mainly in dense breasts [1]. The American College of Radiology developed the Breast Imaging Reporting and Data System (BI-RADS) ultrasound lexicon to provide a common language for reporting and to avoid ambiguity in the interpretation, improving the diagnostic efficiency of the ultrasound [2-4]. The sonographic characteristics are organized into five categories, which are: shape, orientation, margins, echo pattern and posterior acoustic transmission [5,6]. Unfortunately, its diagnostic specificity is poor and generate a significant number of false positive results, increasing biopsied cases [7].
Elastography has been introduced to overcome these limitations and obtain a more accurate characterization of breast lesions. This is a newly developed dynamic technique that uses ultrasound to provide an estimate of tissue stiffness by measuring the degree of distortion under the application of an external force. Like palpation during physical examination, elastography uses tissue deformation or strain, caused by compression and is estimated by precompression and post compression ultrasonic signals. Elastography has proven to be highly specific in the evaluation of lesions. However, strain elastography provided objective data on tissue stiffness via the quantity of tissue displacement [4,7,8].
In this context, errors due to the subjectivity in boundaries definitions and superposition between benign and malignant characteristics are very common during the visual analysis of the specialist. With the advance of digital technology, mainly of the digital image processing - including pattern recognition and artificial intelligence - radiologists have the opportunity to improve the diagnostic accuracy with the aid of computer systems. Computer Aided-Diagnosis (CAD) is a technology which has been implemented with the purpose of providing double reading, working as a second opinion. CAD systems are useful when there is high interobserver variability, absence of trained observers or impossibility of performing double reading with two or more radiologists. Clinical studies have demonstrated that CAD increases sensitivity in the diagnostic of breast cancer [9-11].
This study presents the proposal of investigating the diagnostic ability of a computational system in the characterization of suspicious findings in B-mode ultrasound and breast elastography imaging. The system provides the individual result of each exam, as well as the result of the combination of them, proving to be an innovative classification proposal. We also evaluated the performance of this system in the combined diagnostic with the specialist.

Materials and Methods

Image database:
The local institutional review board approved this study (Protocol No. 53543016.2.0000.0072) and the consent was obtained from all patients. A radiologist with 2 years of experience performed the B-mode ultrasound and Strain Elastography (SE) examinations using a Toshiba Aplio 400 Ultrasound System (Toshiba, Japan) with a 5-10 MHz linear transducer.
The target population was comprised of 83 consecutive female patients, represented by 92 solid lesions. However, we excluded five patients because they presented non-mass lesions on the ultrasound before the percutaneous biopsy confirmation. A total of 83 lesions were included in this study, resulting in 31 malignant and 52 benign lesions. All lesions underwent excisional biopsy, core needle biopsy or fine-needle aspiration biopsy for pathologic diagnosis, used as the gold standard for evaluation of the CAD. The collection of cases was from July to December 2015 during diagnostic breast exams at the Brazilian Institute for Cancer Control (IBCC - São Paulo, SP, Brazil).
The images were deidentified for patient confidentiality. The SE image was superimposed on the corresponding B-mode grayscale image with a color scale. In the color scale, blue indicates soft tissue and red indicates hard tissue. B-mode images were on the right side and elastographic images were on the left side.
Delimitation of the lesion:
We evaluated the diagnosis from manual delimitation and using an automatic segmentation technique.
Three radiologists draw the contour on the B-mode ultrasound images. The first has 16 years of experience in breast imaging, the second has 10 years and the last one was a second-year resident.
In the automatic segmentation, we used the active contour technique based on the Mumford-Shah and level set functions [12]. We also applied a post-processing to remove disconnected pixels and join internal valleys, as described in a previous work [13].
Classification in B-mode ultrasound imaging:
The quantitative features extracted from segmented image can be classified, according to the fifth edition of Breast Imaging Reporting and Data System (BI-RADS), into five categories: shape, orientation, margin, echo pattern and posterior acoustic features. Ten features were de-fined by in order to quantify these BI-RADS features, as presented in [14] Table 1.
Figure 1: Representative image of setting the region of interest for CAD system analysis. The margin of the breast mass was defined manually for analysis (yellow line). The ultrasonographic and elastographic features were automatically analyzed by the CAD system, and a final assessment was visualized.
In order to distinguish benign from malignant lesions, we used the Support Vector Machine (SVM) [16], which is a technique that seeks an optimal hyperplane to separate two classes of samples. The function FITCSVM was applied to create the routine in MATLAB. 10-fold cross-validation was used to evaluate the performance of the classifier. The best performance of the classifier was using the features ADEE, orientation, NumPeaks, entropy and lesion size. The procedure applied to select features and classify the lesions were described in more detail by [17].
Classification in elastography imaging:
In order to measure the amount of hard tissue (i.e., tissues in red) in the lesion, we developed an algorithm to segment red areas and quantify its predominance within the lesion, allowing us to classify it as soft, intermediate, or hard. This algorithm converts the image from RGB to CIELab color space and, after that, we applied Otsu method on the a* channel [18,19].
Table 1: Summary of the features that quantify each BI-RADS category.
Table 2: Classification of the breast lesions after the addition of the elastography results (soft, intermediate and hard) to the BI-RADS lexicon [21].
Table 3: Classification of breast lesions in ultrasonography.
We classified the lesions in elastography images into three categories: (1) soft when the red area is lower than 50% of the total lesion area; (2) intermediate when this value is between 50-75%; and (3) hard when the red area is larger than 75% of the total area. We considered the lesions classified as soft and intermediate as negative and hard as positive cases. We included in the system the possibility of the specialist to change the threshold value. This value defines whether the pixel is red or any other color. The specialist can change the threshold value if he considers that the color distribution is not entirely accurate based on his visual perception [20].
Association between ultrasound and elastography:
Fleury (2015) proposed an association criterion between ultrasound and elastography results in order to provide a single diagnosis. This criterion is presented in Table 2.
To use this classification criterion, we assign percentage values for the tendency to malignity or benignity, since the SVM provides a binary result, where zero defines the lesion as benign and 1 as malignant.
To make it possible to use this classification criterion, we used the values of separation between the hyperplanes. The value of separation ranges between -1 and +1. Thus, the values between +1 and -1 represents the variation from zero to 100%, where zero is for benign and 100% for malignant lesion.
We defined thresholds to represent each BI-RADS category, which are:
- BI-RADS 3: malignancy percentage less than 50%.
- BI-RADS 4a: malignancy percentage between 50 and 64%.
- BI-RADS 4b: malignancy percentage between 64% and 77%.
- BI-RADS 4c: malignancy percentage between 77% and 90%.
- BI-RADS 5: malignancy percentage greater than 90%.
Image review and application of the CAD system:
A radiologist with 16 years of breast imaging experience reviewed the data for analysis. The observer was blinded to clinical information and pathologic results of each mass during image review. After image review by the radiologist, the CAD system was applied to the same image the radiologist used for his analysis. The Region of Interest (ROI) was either automatically or manually drawn along the border of the mass by CAD system. The results of the CAD and final assessments were immediately displayed and recorded for data analysis Figure 1. After being informed of the final assessment made by CAD system, the radiologist gave a final assessment for each breast mass, integrating the results of the CAD system.
Data evaluation and statistical analysis:
Final assessment based on BI-RADS criterion were also divided into 2 groups for statistical analysis: positive assessments consisted of categories 4a to 5, and negative assessments consisted of categories 2 and 3. Diagnostic performance of the radiologist, CAD system and the integration of CAD with the radiologist were analyzed and compared, including sensitivity, specificity, and area under the receiver operating characteristic curve (AUC).

Results

Classification of B-mode ultrasound:
We applied the classifier developed by on our image database and the results are presented in [17] Table 3.
Data from visual analysis are related to the clinical diagnosis provided by Radiologist 1 not using the computational tool to assist in his diagnosis. However, Radiologist 1 was the only one who performed the visual analysis.
Classification of elastography:
For this stage, we performed two experiments. Experiment 1: we investigated the accuracy of the fully automatic classifier, i.e., we quantified the pixels representing the hard tissue by using Otsu method, from the a* channel. In Experiment 2, the specialist could change the threshold value to adjust the color distribution, this procedure can include or remove the amount of tissue defined as red by the automatic threshold Table 4.
Combined diagnostic of elastography and B-mode ultrasound:
For the final diagnosis, we associated the results of both examinations according to the criterion presented in item 2.E
Table 5 shows the sensitivity, specificity and AUC values for the diagnostic combination of the CAD system for B-mode ultrasound and elastography. We evaluated the result from elastography CAD system with and without the radiologist intervention.
Final assessment between the radiologist and the CAD system:
Interobserver variability is inevitable, and it can lead to inconsistent diagnoses among radiologists [22,23]. CAD systems have recently been used to overcome this variability and increase the diagnostic accuracy of breast lesions.
Table 4: Classification of breast lesions in elastography.
Table 5: Result of the diagnostic combination between B-mode ultrasound and elastography.
Table 6: CAD influence on the radiologist’s classification.
The final evaluation consisted of the integration of radiologist and CAD system. For this analysis, B-mode ultrasound and elastography image were presented to the radiologist, who provided the diagnosis based on his visual analysis. Then, the results of the system (individual and with the association of the exams) were shown to him. The radiologist classified the lesion again and we can evaluated the influence of CAD on his diagnosis. The results obtained are shown in Table 6.

Discussion

The analysis of the images, in general, occurs through the visual analysis performed by one, or when possible, more specialists. However, such visual analysis may result in significant inter and intraobserver variability, even when the procedure is performed under the same conditions [24,25]. Some studies report the interobserver variability in the diagnosis of lesions in breast ultrasound and elastography imaging [26-29].
Computer-aided diagnosis systems were recently applied to overcome the variability observer as well as to improve the diagnostic performances. Our CAD system applies a novel classification technique, providing diagnoses for breast lesions found on B-mode ultrasound and elastography images. In addition, it is possible to obtain the combined diagnosis of the exams (US + elastography).
Among the methods of machine learning, SVM is a classifier that has been widely used to distinguish benign from malignant lesions in B-mode ultrasound imaging [14,30], including commercial systems such as, for example, S-detect and B-CAD [23,31,32]. Our study showed a superior result to S-detect - whose AUC was 0.815, while the proposed CAD system was 0.843, considering the manual delineation of the most experienced radiologist.
Elastography provides significant information regarding tissue elasticity and suspicious findings. This information is expressed by color variation during compression and after decompression of ROI. The proposed approach is simple and capable of increasing diagnostic specificity, as presented in some studies and comparable or superior to the results of other systems [33-36]. The system allows parameter adjustment in order to increase the diagnostic accuracy as reported in Experiment 2, where Radiologist 2 was able to improve diagnostic accuracy, obtaining AUC value greater than that Radiologist 1 at 1.02%.
The model developed for classification in elastography images is related to the shape of the lesion (segmented area), the AUC value for the automatic method is lower than the value obtained by radiologists due to the inclusion of surrounding tissues. However, when the result is associated with that obtained in the classification of the B-mode ultrasound image, the AUC value approximates that provided with the manual contour delimitation of Radiologists 1 and 2 and is higher than that of the Resident.
From the data shown in Table 5, we observed that the interference of Radiologist 1 in the threshold value was not significant when this classification is associated with that of the B-mode ultrasound, since the sensitivity, specificity and AUC were not changed. When Radiologist 2 changed threshold value, specificity and AUC increased by 3.22% and 1.25%, respectively. On the other hand, the value of specificity and AUC decreased when the Resident changed the value of the threshold.
We did not include the automatic segmentation in Experiment 2 due to the absence of an observer to perform the change in threshold value. However, the user can change the threshold value if he thinks it is necessary. It is important to highlight the change of this value can modify the result of the classification in the elastography and also in the combined diagnosis.
The visual evaluation of the radiologist with the aid of the CAD system provided a substantial increase in the specificity rate from 55.77% (29/52) to 73.08% (38/52) - addition of 11 true Negative cases (VN) and two False Positives (FP). In relation to sensitivity, the variation was lower, about 3% - increase of two true Positive cases (PV) and one false Negative (VN), resulting in a sensitivity of 90.32% (28/31). Thus, we can affirm that the developed system was able to improve overall diagnosis, increasing the sensitivity and specificity when used in conjunction with the specialist’s clinical evaluation, providing a significant increase in AUC value (14.43%).
The main reasons for the radiologist’s change of opinion after he sees the result of the CAD sys-tem are related to the visualization of the morphology of the lesion after the manual delineation and the distribution/quantification of the hard and soft tissues within the lesion in elastography imaging. This factor may be related to the more precise visualization of the margins and the shape of the lesion when it is seen with the overlap of the contour.
In addition, the behavior of the CAD system was similar to the visual analysis of the radiologist, reaching a higher sensitivity rate in the diagnosis of lesions seen in the B-mode ultrasound and in the association of the exams and, for the elastography, a higher specificity rate.
In conclusion, our initial experience with ultrasound breast elastography showed that it was more specific and more accurate than conventional ultrasound. Combining our system with the experience of the radiologist can improve the specificity and can potentially reduce unnecessary breast biopsies.

Acknowledgement

This work was supported by the São Paulo Research Foundation (FAPESP) grant #2012/24006-5. The content is solely the responsibility of the authors and does not necessarily represent the official views of FAPESP.

References