Journal of Cancer Sciences
Download PDF
Research Article
Breast Cancer Classification: A CAD System for a Combined Use of Elastography and B-Mode Sonography
Marcomini KD1*, Fleury EFC2 and H. Schiabel2
1Department of Electrical and Computer Engineering, University of São Paulo, Brazil
2Department of Electrical and Computer Engineering, University of So Paulo,Brazil
*Address for Correspondence: Marcomini KD, Department of Electrical and Computer Engineering, University of São Paulo, 400 Avenida Trabalhador São-Carlense, 13566-590 São Carlos, SP, Brazil
Submission: 05 August, 2020;
Accepted: 12 September, 2020;
Published: 20 November, 2020
Copyright: © 2020 Marcomini KD, et al. This is an open access article
distributed under the Creative Commons Attribution License, which
permits unrestricted use, distribution, and reproduction in any medium,
provided the original work is properly cited.
Abstract
Purpose: The aim of this study was to evaluate and compare the
diagnostic performance of elastography, B-mode Ultrasound (US), and
a combination of elastography and B-mode US for the differentiation
between benign and malignant lesions.
Methods: A prospective study was carried out from July to
December 2015, which 87 patients with 83 lesions were examined
with conventional B-mode ultrasound and strain elastography. All
the lesions had been proven by biopsy, resulting in 31 malignant and
52 benign lesions. A radiologist with 16 years of experience classified
visually these cases. We also used a CAD sys-tem to classify the lesions
classified visually by the most experienced radiologist and using a CAD
system. The data obtained were compared with the results provided
by another radiologist and a resident with 2 years of experience.
Sensitivity, specificity and AUC for the three observers using the CAD
system were calculated.
Results: The developed CADx system provided a diagnostic
concordance in the classification of breast lesions from the different
ways of contour determination (manual and automatic), allowing
to reduce the diagnostic variability. In addition, the CADx system
showed superior results to the visual analysis of the radiologist. When
the radiologist associated both examinations (B-mode ultrasound and
elastography), his visual analysis provided 87.1%, 55.8% and 0.714 of
sensitivity, specificity and AUC, respectively. When we considered
the result provided by the association between B-mode ultrasound
and elastography images, the CADx system provided a comparative
increase of about 7% of sensitivity and 17.2% of specificity, using the
contour delimited by the most experienced radiologist. In addition,
a positive influence was observed in the use of the computational
tool by radiologists, since, on average, their sensitivity and specificity
indexes also increased in relation to the conventional analysis, from
87.1% and 55.8% to 90.3% and 73.1%, respectively.
Conclusion: Thus, it can be concluded that the developed CADx
system performed well in distinguishing benign from malignant lesions
for both B-mode ultrasound and elastography. The AUC obtained
was higher than the radiologist’s visual analysis in most of the cases
analyzed.
Keywords
Breast cancer; Elastography imaging; Computer-Aided Diagnosis; Color map; Interobserver agreement
Introduction
Breast Ultrasound (US) is an important complementary technique
for screening and has been proven to be useful in differentiating benign
from malignant masses, mainly in dense breasts [1]. The American
College of Radiology developed the Breast Imaging Reporting and
Data System (BI-RADS) ultrasound lexicon to provide a common
language for reporting and to avoid ambiguity in the interpretation,
improving the diagnostic efficiency of the ultrasound [2-4]. The
sonographic characteristics are organized into five categories, which are: shape, orientation, margins, echo pattern and posterior acoustic transmission [5,6]. Unfortunately, its diagnostic specificity is poor and generate a significant number of false positive results, increasing
biopsied cases [7].
Elastography has been introduced to overcome these limitations
and obtain a more accurate characterization of breast lesions. This is
a newly developed dynamic technique that uses ultrasound to provide
an estimate of tissue stiffness by measuring the degree of distortion
under the application of an external force. Like palpation during
physical examination, elastography uses tissue deformation or strain,
caused by compression and is estimated by precompression and post
compression ultrasonic signals. Elastography has proven to be highly
specific in the evaluation of lesions. However, strain elastography
provided objective data on tissue stiffness via the quantity of tissue
displacement [4,7,8].
In this context, errors due to the subjectivity in boundaries
definitions and superposition between benign and malignant
characteristics are very common during the visual analysis of the
specialist. With the advance of digital technology, mainly of the
digital image processing - including pattern recognition and artificial
intelligence - radiologists have the opportunity to improve the
diagnostic accuracy with the aid of computer systems. Computer
Aided-Diagnosis (CAD) is a technology which has been implemented
with the purpose of providing double reading, working as a second
opinion. CAD systems are useful when there is high interobserver
variability, absence of trained observers or impossibility of performing
double reading with two or more radiologists. Clinical studies have
demonstrated that CAD increases sensitivity in the diagnostic of
breast cancer [9-11].
This study presents the proposal of investigating the diagnostic
ability of a computational system in the characterization of suspicious
findings in B-mode ultrasound and breast elastography imaging. The
system provides the individual result of each exam, as well as the result
of the combination of them, proving to be an innovative classification
proposal. We also evaluated the performance of this system in the
combined diagnostic with the specialist.
Materials and Methods
Image database:
The local institutional review board approved this study (Protocol
No. 53543016.2.0000.0072) and the consent was obtained from all
patients. A radiologist with 2 years of experience performed the
B-mode ultrasound and Strain Elastography (SE) examinations using
a Toshiba Aplio 400 Ultrasound System (Toshiba, Japan) with a 5-10
MHz linear transducer.The target population was comprised of 83 consecutive female
patients, represented by 92 solid lesions. However, we excluded five
patients because they presented non-mass lesions on the ultrasound
before the percutaneous biopsy confirmation. A total of 83 lesions
were included in this study, resulting in 31 malignant and 52 benign
lesions. All lesions underwent excisional biopsy, core needle biopsy
or fine-needle aspiration biopsy for pathologic diagnosis, used as the
gold standard for evaluation of the CAD. The collection of cases was
from July to December 2015 during diagnostic breast exams at the
Brazilian Institute for Cancer Control (IBCC - São Paulo, SP, Brazil).
The images were deidentified for patient confidentiality. The SE
image was superimposed on the corresponding B-mode grayscale
image with a color scale. In the color scale, blue indicates soft tissue
and red indicates hard tissue. B-mode images were on the right side
and elastographic images were on the left side.
Delimitation of the lesion:
We evaluated the diagnosis from manual delimitation and using
an automatic segmentation technique.Three radiologists draw the contour on the B-mode ultrasound
images. The first has 16 years of experience in breast imaging, the
second has 10 years and the last one was a second-year resident.
In the automatic segmentation, we used the active contour
technique based on the Mumford-Shah and level set functions [12].
We also applied a post-processing to remove disconnected pixels and
join internal valleys, as described in a previous work [13].
Classification in B-mode ultrasound imaging:
The quantitative features extracted from segmented image can be
classified, according to the fifth edition of Breast Imaging Reporting and Data System (BI-RADS), into five categories: shape, orientation,
margin, echo pattern and posterior acoustic features. Ten features
were de-fined by in order to quantify these BI-RADS features, as
presented in [14] Table 1.
Figure 1: Representative image of setting the region of interest for CAD
system analysis. The margin of the breast mass was defined manually for
analysis (yellow line). The ultrasonographic and elastographic features were
automatically analyzed by the CAD system, and a final assessment was
visualized.
In order to distinguish benign from malignant lesions, we used
the Support Vector Machine (SVM) [16], which is a technique that
seeks an optimal hyperplane to separate two classes of samples. The
function FITCSVM was applied to create the routine in MATLAB.
10-fold cross-validation was used to evaluate the performance of
the classifier. The best performance of the classifier was using the
features ADEE, orientation, NumPeaks, entropy and lesion size.
The procedure applied to select features and classify the lesions were
described in more detail by [17].
Classification in elastography imaging:
In order to measure the amount of hard tissue (i.e., tissues in red)
in the lesion, we developed an algorithm to segment red areas and
quantify its predominance within the lesion, allowing us to classify it
as soft, intermediate, or hard. This algorithm converts the image from
RGB to CIELab color space and, after that, we applied Otsu method on the a* channel [18,19].
Table 2: Classification of the breast lesions after the addition of the elastography results (soft, intermediate and hard) to the BI-RADS lexicon [21].
We classified the lesions in elastography images into three categories: (1) soft when the red area is lower than 50% of the total
lesion area; (2) intermediate when this value is between 50-75%; and
(3) hard when the red area is larger than 75% of the total area. We
considered the lesions classified as soft and intermediate as negative
and hard as positive cases. We included in the system the possibility
of the specialist to change the threshold value. This value defines
whether the pixel is red or any other color. The specialist can change
the threshold value if he considers that the color distribution is not
entirely accurate based on his visual perception [20].
Association between ultrasound and elastography:
Fleury (2015) proposed an association criterion between
ultrasound and elastography results in order to provide a single
diagnosis. This criterion is presented in Table 2.To use this classification criterion, we assign percentage values
for the tendency to malignity or benignity, since the SVM provides
a binary result, where zero defines the lesion as benign and 1 as
malignant.
To make it possible to use this classification criterion, we used the values of separation between the hyperplanes. The value of separation ranges between -1 and +1. Thus, the values between +1 and -1
represents the variation from zero to 100%, where zero is for benign and 100% for malignant lesion.
We defined thresholds to represent each BI-RADS category, which are:
- BI-RADS 3: malignancy percentage less than 50%.
- BI-RADS 4a: malignancy percentage between 50 and 64%.
- BI-RADS 4b: malignancy percentage between 64% and 77%.
- BI-RADS 4c: malignancy percentage between 77% and 90%.
- BI-RADS 5: malignancy percentage greater than 90%.
Image review and application of the CAD system:
A radiologist with 16 years of breast imaging experience reviewed
the data for analysis. The observer was blinded to clinical information
and pathologic results of each mass during image review. After
image review by the radiologist, the CAD system was applied to
the same image the radiologist used for his analysis. The Region of
Interest (ROI) was either automatically or manually drawn along
the border of the mass by CAD system. The results of the CAD and
final assessments were immediately displayed and recorded for data
analysis Figure 1. After being informed of the final assessment made
by CAD system, the radiologist gave a final assessment for each breast
mass, integrating the results of the CAD system.Data evaluation and statistical analysis:
Final assessment based on BI-RADS criterion were also divided
into 2 groups for statistical analysis: positive assessments consisted
of categories 4a to 5, and negative assessments consisted of categories
2 and 3. Diagnostic performance of the radiologist, CAD system
and the integration of CAD with the radiologist were analyzed and
compared, including sensitivity, specificity, and area under the
receiver operating characteristic curve (AUC).Results
Classification of B-mode ultrasound:
We applied the classifier developed by on our image database and
the results are presented in [17] Table 3.Data from visual analysis are related to the clinical diagnosis
provided by Radiologist 1 not using the computational tool to assist
in his diagnosis. However, Radiologist 1 was the only one who
performed the visual analysis.
Classification of elastography:
For this stage, we performed two experiments. Experiment 1:
we investigated the accuracy of the fully automatic classifier, i.e.,
we quantified the pixels representing the hard tissue by using Otsu
method, from the a* channel. In Experiment 2, the specialist could
change the threshold value to adjust the color distribution, this
procedure can include or remove the amount of tissue defined as red
by the automatic threshold Table 4.Combined diagnostic of elastography and B-mode ultrasound:
For the final diagnosis, we associated the results of both
examinations according to the criterion presented in item 2.ETable 5 shows the sensitivity, specificity and AUC values for the
diagnostic combination of the CAD system for B-mode ultrasound
and elastography. We evaluated the result from elastography CAD
system with and without the radiologist intervention.
Final assessment between the radiologist and the CAD system:
Interobserver variability is inevitable, and it can lead to
inconsistent diagnoses among radiologists [22,23]. CAD systems
have recently been used to overcome this variability and increase the
diagnostic accuracy of breast lesions.The final evaluation consisted of the integration of radiologist and
CAD system. For this analysis, B-mode ultrasound and elastography
image were presented to the radiologist, who provided the diagnosis
based on his visual analysis. Then, the results of the system (individual
and with the association of the exams) were shown to him. The
radiologist classified the lesion again and we can evaluated the
influence of CAD on his diagnosis. The results obtained are shown
in Table 6.
Discussion
The analysis of the images, in general, occurs through the visual
analysis performed by one, or when possible, more specialists.
However, such visual analysis may result in significant inter and intraobserver
variability, even when the procedure is performed under
the same conditions [24,25]. Some studies report the interobserver
variability in the diagnosis of lesions in breast ultrasound and
elastography imaging [26-29].
Computer-aided diagnosis systems were recently applied to
overcome the variability observer as well as to improve the diagnostic
performances. Our CAD system applies a novel classification
technique, providing diagnoses for breast lesions found on B-mode
ultrasound and elastography images. In addition, it is possible to
obtain the combined diagnosis of the exams (US + elastography).
Among the methods of machine learning, SVM is a classifier
that has been widely used to distinguish benign from malignant
lesions in B-mode ultrasound imaging [14,30], including commercial systems such as, for example, S-detect and B-CAD [23,31,32]. Our study showed a superior result to S-detect - whose AUC was 0.815, while the proposed CAD system was 0.843, considering the manual delineation of the most experienced radiologist.
Elastography provides significant information regarding tissue
elasticity and suspicious findings. This information is expressed by
color variation during compression and after decompression of ROI.
The proposed approach is simple and capable of increasing diagnostic
specificity, as presented in some studies and comparable or superior
to the results of other systems [33-36]. The system allows parameter
adjustment in order to increase the diagnostic accuracy as reported
in Experiment 2, where Radiologist 2 was able to improve diagnostic
accuracy, obtaining AUC value greater than that Radiologist 1 at
1.02%.
The model developed for classification in elastography images is
related to the shape of the lesion (segmented area), the AUC value for
the automatic method is lower than the value obtained by radiologists
due to the inclusion of surrounding tissues. However, when the result
is associated with that obtained in the classification of the B-mode
ultrasound image, the AUC value approximates that provided with
the manual contour delimitation of Radiologists 1 and 2 and is higher
than that of the Resident.
From the data shown in Table 5, we observed that the interference
of Radiologist 1 in the threshold value was not significant when
this classification is associated with that of the B-mode ultrasound,
since the sensitivity, specificity and AUC were not changed. When
Radiologist 2 changed threshold value, specificity and AUC increased
by 3.22% and 1.25%, respectively. On the other hand, the value of
specificity and AUC decreased when the Resident changed the value
of the threshold.
We did not include the automatic segmentation in Experiment 2
due to the absence of an observer to perform the change in threshold
value. However, the user can change the threshold value if he thinks
it is necessary. It is important to highlight the change of this value can modify the result of the classification in the elastography and also in
the combined diagnosis.
The visual evaluation of the radiologist with the aid of the CAD
system provided a substantial increase in the specificity rate from
55.77% (29/52) to 73.08% (38/52) - addition of 11 true Negative
cases (VN) and two False Positives (FP). In relation to sensitivity, the
variation was lower, about 3% - increase of two true Positive cases
(PV) and one false Negative (VN), resulting in a sensitivity of 90.32%
(28/31). Thus, we can affirm that the developed system was able to
improve overall diagnosis, increasing the sensitivity and specificity
when used in conjunction with the specialist’s clinical evaluation,
providing a significant increase in AUC value (14.43%).
The main reasons for the radiologist’s change of opinion after he
sees the result of the CAD sys-tem are related to the visualization of
the morphology of the lesion after the manual delineation and the
distribution/quantification of the hard and soft tissues within the
lesion in elastography imaging. This factor may be related to the more
precise visualization of the margins and the shape of the lesion when
it is seen with the overlap of the contour.
In addition, the behavior of the CAD system was similar to the
visual analysis of the radiologist, reaching a higher sensitivity rate in
the diagnosis of lesions seen in the B-mode ultrasound and in the
association of the exams and, for the elastography, a higher specificity
rate.
In conclusion, our initial experience with ultrasound breast
elastography showed that it was more specific and more accurate than
conventional ultrasound. Combining our system with the experience
of the radiologist can improve the specificity and can potentially
reduce unnecessary breast biopsies.
Acknowledgement
This work was supported by the São Paulo Research
Foundation (FAPESP) grant #2012/24006-5. The content is solely
the responsibility of the authors and does not necessarily represent
the official views of FAPESP.